TDK Lambda | Z+Low Voltage User Manual
7.12.2 Instrument Subsystem The subsystem programs one or more power supplies.
SCPI Command Syntax GEN Command Syntax
INSTrument:COUPle
None
Parameters
ALL|NONE
INSTrument:NSELect
SCPI Command Syntax GEN Command Syntax
INSTrument:NSELect
ADR n (single address) INSTrument:NSELect 6 INSTrument:NSELect?
Examples
Query Syntax
7.12.3 Voltage Subsystem VOLTage Sets the output voltage value in Volts. The range of voltage values are described in Table 7-5. The maximum number of characters is 12.
SCPI Command Syntax GEN Command Syntax
[SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPLitude]
PV < NRf+> PV?
Default Suffix
V 0
*RST Value Examples
:VOLT 500 MV
VOLT:LEV 234.56789
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]? VOLTage? MAX VOLTage? MIN
Query Syntax
Returned Parameters
VOLTage:MODE This command selects FIX, LIST, WAVE subsystems control over the power supply output voltage.
SCPI Command Syntax GEN Command Syntax
[SOURce]:VOLTage:MODE
None
Parameters *RST Value Examples Query Syntax
NONE | FIXed | LIST | WAVE
NONE
VOLT:MODE LIST VOLT:MODE FIX
[SOURce]:VOLTage:MODE? NONE | FIX | LIST | WAVE
Returned Parameters
102
Made with FlippingBook Annual report maker