TDK Lambda | Z+High Voltage User Manual
7.12.7 INITiate Subsystem INITiate Enables the trigger subsystem. If a trigger circuit is not enabled, all trigger commands are ignored.
SCPI Command Syntax GEN Command Syntax
INITiate[:IMMediate]
None None
Parameters Examples Query Syntax
INIT:IMM
None
INITiate:CONTinuous • INIT:CONT 0 - Enables the trigger subsystem only for a single trigger action. The subsystem must be enabled prior to each subsequent trigger action. • INIT:CONT 1 - Trigger system is continuously enabled and INIT is redundant.
SCPI Command Syntax
INITiate:CONTinuous
Parameters Examples Query Syntax
0|OFF 1|ON
INIT:CONT 1 INIT:CONT ON
INITiate:CONTinuous?
Returned Parameters
0 |1
NOTE: When power supply is in INIT:CONTONmode, it’s impossible to change any programming parameter. Display shows ” Err ” . Send ABORt commands to allow parameter change.
7.12.8 LIST Subsystem This subsystem receives parameters to create a step waveform of the output voltage or current.
NOTE: All list subsystem commands (as well as CURR:MODE LIST and VOLT:MODE LIST) are implied ABORt commands.
LIST:COUNt Sets the number of times that the list is executed before it is completed. The command accepts parameters in the range 1 through 9999, but any number greater than 9999 is interpreted as INFinity. Use INF if you wish to execute a list indefinitely.
SCPI Command Syntax
[SOURce]:LIST:COUNt
Parameters
1 to 9999 | INFinity
*RST
1
Examples
LIST:COUN 3 LIST:COUN INF
Query Syntax
[SOURce]:LIST:COUNt? (if count>9999 the response is INF)
Returned Parameters
107